Calibration · adjustment · verification Calibration, verification and adjustment: the difference

“The micrometer needs calibrating” can mean three different jobs. Calibration compares, adjustment changes, verification decides – and the order in which you do them decides what you can prove afterwards.

Last updated: November 6, 2026 · Reviewed by irot

  • Free for up to 20 devices
  • No payment details
  • Interface in English and German
Calibration
Recording a calibration in Kalibrat

Three words, three actions

“The micrometer needs calibrating” can mean three different jobs. They differ in what happens to the gauge, and therefore in what you can prove afterwards.

Calibration
Comparing what the gauge shows with a better, traceable reference and writing down the difference – at several points, with the uncertainty of the comparison. The gauge is not changed. Example: a 25 mm micrometer is measured against gauge blocks and reads 0.003 mm high at 12.5 mm.
Adjustment
Setting the gauge so that its error becomes smaller – zeroing a dial, turning the setting nut, re-trimming a scale. This does change the gauge. Example: you turn the micrometer's sleeve so that it reads zero when closed.
Verification
Checking the result against a limit and deciding whether the gauge is fit for its job. Example: the 0.003 mm error is compared with the tolerance you work to, and the answer is “fine for this job” or “not fine”.

Calibration on its own never says “pass” or “fail”. It gives you the numbers; the decision comes from comparing them with a limit set by the manufacturer, a standard or your own requirements. In everyday shop talk, “verification” is also used for a quick check without a full calibration – see below.

Why the order matters

If you adjust a gauge and only then measure it, you have lost the one piece of information that matters most: how the gauge was measuring before you touched it. That reading is called as found; the reading after the adjustment is as left.

  1. Calibrate – record the as-found state.
  2. Decide – is the error inside the limit, allowing for the uncertainty?
  3. If needed, adjust, then calibrate again to record the as-left state.
  4. If the as-found state was outside the limit, look back: which parts were measured with this gauge since the last good check?

Step four is what auditors ask about. A certificate that shows only the state after adjustment cannot answer it.

Verification in the workshop

Between two calibrations a gauge can be dropped, knocked or simply worn. A cheap interim check catches that: measure a gauge block or a ring gauge with the caliper before the shift, or once a week, and write down the reading. If it is inside the limit you carry on; if not, the caliper is taken out of use until it has been looked at. This is not a calibration – there is no traceable reference for every point and no uncertainty budget – but it shows that the gauge is still behaving between the dates on the certificate.

Legal metrology: a different thing again

The German word Eichung is a common source of confusion. Its English counterpart is verification under legal metrology (weights and measures law): a check carried out by a public authority, or a body it has approved or notified, on instruments used for trade or for other regulated purposes. Think of scales at a market stall, fuel pumps, and electricity, gas or water meters. Such instruments carry a mark showing they have passed, and the legal period between checks is set by law, not by you.

A workshop micrometer or caliper is normally not in that group. It is calibrated or verified in the quality-management sense, because you need to show that it is fit for the measurements you make. Which rules apply depends on how the instrument is used and where, so if you are unsure, ask your local weights-and-measures authority.

What goes in the record

  • Which device and which date.
  • The result – pass, pass with a restriction, or fail – and what was done about it: released, restricted, adjusted, blocked, sent for repair, withdrawn.
  • Who did it – you, or an outside laboratory, with the certificate number and, if the laboratory is accredited, its accreditation number.
  • The reference used, so that the chain back to a national standard can be followed.
  • Conditions that matter, such as room temperature for length measurement.
  • The assessment of earlier results whenever a gauge turns out to be out of tolerance.

Adjustment deserves its own line in the history. After an adjustment the gauge starts again, and the drift you saw before no longer predicts how it will behave next.

Calibration, adjustment and verification in Kalibrat

Kalibrat keeps one entry per check on each device. It records the result – passed, passed with a restriction, or failed – and the action taken: released, released with a restriction, adjusted or repaired and released, blocked, sent for repair or recalibration, or withdrawn from use. For checks done by an outside laboratory it also holds the certificate number and, optionally, the accreditation number, and for measuring equipment you can name the reference gauge you checked against and the room temperature. If a device is found out of order, Kalibrat will not save the entry until you have written down your assessment of the earlier measurements. A wrong entry is cancelled with a stated reason rather than silently changed. When a device was adjusted or repaired, its interval review starts again from that entry. Free for up to 20 devices – try it with your own gauges.

FAQ

Frequently asked questions

Is calibration the same as adjustment?

No. Calibration measures and records the error without changing the instrument. Adjustment changes the instrument to reduce the error. A calibration before and after the adjustment gives the as-found and as-left states.

What is the difference between calibration and verification?

Calibration gives you numbers: the deviation at several points with its uncertainty. Verification compares those numbers with a limit and decides whether the instrument is fit for its job.

Do I have to calibrate or verify my gauges?

If measurements with them show that a product meets its requirements, a quality system such as ISO 9001 expects calibration or verification against traceable standards at specified intervals – your company sets the intervals.

What does the German word “Eichung” mean?

Verification under legal metrology: a check by a public authority or an approved body on instruments used for trade or other regulated purposes, such as market scales or fuel pumps. A workshop micrometer is normally not covered.

One record per check, with the result and the action – free up to 20 devices

Free for up to 20 devices, for as long as you like, with no payment details. Set up in three minutes – e-mail, your company, done. Your spreadsheet comes along.

Start free See pricing