Measurement uncertainty · GUM budget Measurement uncertainty calculator: the budget in four lines

Enter the uncertainty of the reference, the resolution, your readings and the temperature – uc, U and the share of each contribution update as you type. No sign-up, and nothing leaves your browser.

Last updated: November 6, 2026 · Reviewed by irot

  • Free for up to 20 devices
  • No payment details
  • Interface in English and German
Uncertainty
Measurement uncertainty in Kalibrat

The calculator

Enter all values in the same unit as your measurement (for example mm) – only the expansion coefficient in 10⁻⁶/K and the temperature difference in K. Empty fields do not count.

How the budget is calculated

Reference: Uref / 2
The calibration certificate states the expanded uncertainty with k = 2; the standard uncertainty is half of it.
Resolution: d / (2·√3)
The true value lies anywhere within half a digit step – equally likely, so a rectangular distribution.
Repeatability: s
The standard deviation of repeated readings on the same reference.
Temperature: L·Δα·ΔT / √3
The length error when the item and the reference expand differently and the measurement is not made at 20 °C – taken as rectangular.

The contributions are added in quadrature: uc = √(Σ ui²). The expanded uncertainty is U = 2·uc, rounded up to two significant digits – up, because an uncertainty stated too small makes the decision look safer than the measurement supports. This is the usual form for calibrating measuring instruments; a full budget may need further contributions (fixture, operator, long-term drift).

What U has to do with the tolerance

When a reading is close to a limit, the uncertainty decides whether it is safely inside, safely outside or in between. A decision rule says how to treat that zone. As a rule of thumb: below 10 % of the tolerance U hardly matters, above 25 % the instrument is too imprecise for that tolerance.

Same calculation as in the Kalibrat application (Metrology module), where the budget sits with each calibration and the uncertainty of the reference is pre-filled from its latest calibration. Based on the GUM (JCGM 100) as commonly applied to measuring instruments; this budget does not replace a complete uncertainty evaluation.

FAQ

Frequently asked questions

What is the difference between u_c and U?

u_c is the combined standard uncertainty, U the expanded uncertainty. With k = 2, U covers about 95 % of the values that could reasonably be attributed to the measurand, assuming a normal distribution.

Why is the resolution divided by 2·√3?

A reading with digit step d can lie anywhere within ± d/2 with equal probability. The standard deviation of such a rectangular distribution is the half-width divided by √3.

Which contribution is usually the largest?

For hand-held instruments often repeatability or resolution, for fine measurements temperature. The share column shows where improving pays off.

Is my data uploaded?

No. The calculation runs entirely in your browser.