Gauge capability · type 1 study Cg and Cgk calculator: is your gauge capable?
Enter the reference value, the tolerance and your readings – Cg, Cgk and the bias update as you type. No sign-up, and the readings never leave your browser.
Last updated: November 6, 2026 · Reviewed by irot
- Free for up to 20 devices
- No payment details
- Interface in English and German

The calculator
The usual study is 25 readings (some customers ask for 50) of a reference with a known value, by one appraiser, under production conditions. You can paste the values from Excel – one column, one value per line.
What Cg and Cgk tell you
- Cg – the spread
- How much room the repeatability of the gauge takes up in a share of the tolerance: Cg = share · T / (6 s). With the usual share of 20 %, Cg = 1.33 means six standard deviations fill at most 15 % of the tolerance.
- Cgk – spread and bias
- Like Cg, but the bias from the reference value is subtracted: Cgk = (share/2 · T − |x̄ − xref|) / (3 s). A gauge can repeat nicely and still be consistently off – if you only look at Cg, you would call it capable.
- Is the bias significant?
- A t-test tells you whether the deviation from the reference stands out from the noise (p below 0.05). If it does, check zero point, reference and fixture before you blame the instrument.
The gauge is capable if both Cg and Cgk reach at least 1.33 – a common requirement in the automotive supply chain. The share (usually 20 %, sometimes 15 %) and the limit are set by your customer; the calculator uses what you enter.
How to run the study
- A reference with a known value, ideally in the middle of the tolerance, calibrated with traceability.
- One appraiser, one fixture, the same conditions as in production.
- Re-seat the reference every time. Pressing the button twenty-five times measures the electronics, not the measurement system.
- Do not delete anything. An outlier is information about the measurement system.
Type 1, 2 or 3?
The type 1 study comes first: it checks the gauge on its own. Then the Gauge R&R study (type 2) looks at the appraisers on real parts, and type 3 at the measurement system without appraiser influence.
Same calculation as in the Kalibrat application (IATF 16949 module); an automated test runs both versions over the same data. Bosch booklet 10 and the AIAG MSA manual are sold by their publishers and were not checked word for word here; the figures are the widely used requirements, and your customer's requirement is binding.
Frequently asked questions
What is a good Cgk value?
Commonly both Cg and Cgk must reach at least 1.33, based on 20 % of the tolerance. These values come from automotive guidelines and customer requirements, not from a law – your customer's requirement is binding.
Why is Cgk lower than Cg?
Because Cgk also subtracts the bias from the reference value. If both are equal, the mean sits exactly on the reference value.
How many readings does a type 1 study need?
25 are usual, some customers require 50. The calculator also works with fewer but says that the indices rest on a narrower basis.
Is my data uploaded?
No. The calculation runs entirely in your browser; nothing is sent to a server.
Keep capability studies with the gauges they belong to