MSA · Cg, Cgk, %GRR, ndc Measurement system analysis explained
A calibrated gauge can still be useless for a tight tolerance. MSA asks whether the whole measuring process is good enough for one characteristic – three questions, three studies, and limits that are customary rather than law.
Last updated: November 6, 2026 · Reviewed by irot
- Free for up to 20 devices
- No payment details
- Interface in English and German

Why calibration alone is not enough
A calibration certificate answers one question: does this gauge read correctly? It says nothing about whether the gauge is good enough for the feature you are measuring. A micrometer that is perfectly calibrated is excellent for a tolerance of ±0.1 mm and hopeless for a fit with a tolerance of 21 µm. Measurement system analysis (MSA) closes that gap. It looks at the whole measurement system – gauge, appraiser, fixture, method and environment – for one characteristic with its tolerance.
The three questions MSA answers
- Is the gauge capable? Type 1, Cg and Cgk
- One appraiser measures a reference with a known value many times, usually 25. Cg compares the spread of the readings with the tolerance. Cgk also counts the bias – how far the average is from the reference. A gauge that repeats nicely but reads consistently off shows a good Cg and a poor Cgk. Try it with the Cg and Cgk calculator.
- How much of the tolerance does measuring eat? Gage R&R and %GRR
- Several appraisers measure several real parts, each more than once. The analysis splits the variation into repeatability (same gauge, same appraiser), reproducibility (different appraisers) and the parts themselves. %GRR is the share that comes from the measurement system. The Gauge R&R calculator uses analysis of variance and also gives you ndc.
- Can it tell parts apart? ndc
- The number of distinct categories is how many groups the system can still separate within the spread of your parts. With too few, you can sort good from bad at best – you cannot measure, and a control chart built on these values says little.
Which study when
- New gauge
- Start with a type 1 study to see whether the gauge itself is capable, then a Gauge R&R study once real parts and appraisers are involved.
- New or tighter tolerance
- Repeat the studies. The result belongs to a tolerance; if the tolerance halves, the old result no longer answers the question.
- After a repair, a new fixture or a changed method
- The measurement system is no longer the one you studied. Run at least a type 1 study again, and a Gauge R&R if the fixture or the instruction changed.
- Periodically
- Many companies repeat the studies at a fixed interval or when the data start to drift. There is no single correct interval – set one that fits the risk of the characteristic and write it down.
- Automatic measuring equipment
- Where a machine measures and the appraiser has no influence, the type 3 study works like Gauge R&R but leaves out the appraiser.
The usual limits
These are customary acceptance criteria. Your customer may set others, and their requirement is the one that counts.
- Cg and Cgk: at least 1.33, normally with the spread taken as 20 % of the tolerance (some customers ask for 15 %).
- %GRR: under 10 % is generally seen as good, 10 to 30 % as conditionally acceptable and over 30 % as not acceptable.
- ndc: at least 5.
The middle band is a matter of judgement. A %GRR of 20 % may be fine for an uncritical dimension and unacceptable for a safety feature – write down why you accepted it.
Where MSA is expected
ISO 9001 asks for measuring equipment that is traceable and suitable for its purpose, but it does not prescribe a particular statistical study. In the automotive supply chain it is stricter: IATF 16949 expects statistical studies for the measurement systems named in the control plan, usually following the AIAG MSA manual or the German VDA volume 5. The rule of thumb is: not every gauge in the building, but every measurement system that decides whether a part is good or bad. Check what your customer's quality requirements actually say.
Common mistakes
- Parts from one batch. They are too alike, so ndc and %GRR against total variation look terrible while the gauge is fine. Choose parts that cover the tolerance range.
- No characteristic, no tolerance. “MSA for micrometer PM-0042” proves nothing. “Bore Ø 30 H7 on drawing 4711” does.
- Appraisers who can see the part numbers or measure in the same order every time.
- Deleting outliers. An outlier is information about the system, usually the most useful kind.
- Too few readings. Four values still produce a Cgk – one that tells you nothing.
- Mixing up MSA and calibration. One does not replace the other; a measurement system needs both.
MSA in Kalibrat
With the IATF 16949 add-on module, you create a study on the gauge's own record. You choose type 1 (Cg and Cgk), type 2 (Gauge R&R) or type 3 (without appraiser), and you enter the characteristic and its tolerance. Kalibrat calculates Cg and Cgk, %GRR with the reference it used, and ndc, and shows whether the result reaches the usual limits. It warns when the study is smaller than usual. Studies are listed on the device, a cancelled study stays visible as withdrawn, and a report shows for each gauge and characteristic the latest completed study – and which gauges have none yet. The device register itself is free for up to 20 devices – try it with your own gauges.
Frequently asked questions
What is the difference between Cgk and Gage R&R?
Cg and Cgk come from one appraiser measuring a reference many times: they show whether the gauge itself is capable, including its bias. Gage R&R uses several appraisers and real parts and shows how much of the tolerance the whole measurement process uses.
What is a good %GRR?
Customary limits: under 10 % good, 10 to 30 % conditionally acceptable, over 30 % not acceptable. Your customer may set others.
What does ndc mean?
The number of distinct categories: how many groups the measurement system can still separate within the spread of the parts. Customary minimum: 5.
Do I need MSA for ISO 9001?
ISO 9001 does not prescribe a statistical study; it asks for suitable, traceable equipment. IATF 16949 expects studies for the measurement systems in the control plan.
Studies on the gauge's own record – the register is free up to 20 devices
Free for up to 20 devices, for as long as you like, with no payment details. Set up in three minutes – e-mail, your company, done. Your spreadsheet comes along.